Tenney Tenneystrat Altitude Chambers

Tenney’s Tenneystrat Altitude Chambers simulate altitudes up to 200,000 feet. With an overall temperature range of -70°C to +177°C, humidity capabilities of 20% to 98% RH, and workspace sizes ranging from 5 to 64 cubic feet, this highly developed line of altitude chambers offers one of the broadest selections in the industry.

We are confident that our solid altitude chamber construction and precise operating systems will provide the exacting performance and a high degree of reliability for your altitude testing needs.

Altitude Chamber Features

  • Continuously welded, stainless steel, vapor-tight interior liners
  • Control tolerance is ±0.3°C and ±2% RH typical after stabilization
  • Linear pressure transducer to measure altitude
  • Non-corroding humidity system
  • Low-water protection on humidity system
  • Vertical-down recirculating conditioning stream

Altitude Chamber Notes

  • Humidity capability: 20% to 98% RH in the temperature range of +20°C to +85°C as limited by a 3 degree dewpoint.
  • Humidity not achieved simultaneously with altitude.
  • Add suffix "R" to model number for humidity.
  • Vacuum performance: All chambers achieve altitude up to 100,000 feet (8.2 mm Hg) in 35 minutes. Optional features extend altitude capabilities to 150,000 feet (1 mm Hg) and 200,000 feet (0.17 mm Hg).
  • Combined temperature and altitude to 60,000 feet maximum.
  • Test data based on a 24°C ambient, sea level, 60 Hz. Performance slightly reduced on 50 Hz. Power requirements based on standard vacuum pump.

 

Tenney Tenneystrat Altitude Chambers Specifications - Overall dimensions in inches/centimeters

Model T5ST10ST15ST27ST36ST48ST64ST
System Horsepower Size46156156152061520152030152030
WorkspaceW21/5324/6124/6136/9136/9148/12248/122
D21/5324/6136/9136/9148/12236/9148/122
H48/12230/7630/7636/9136/9148/12248/122
ExteriorW25/5151/13052/13063/16063/16077/19677/196
D63/15299/203111/236111/241123/272123-1/2 / 281135-1/2 / 312
H78/18384/21084/21390/22990/22991/23191/231
Heater Capacity246881616
HumdifierWatts123
GPH1.051.01.5
AMPS @ 230V, 333(1)4881528563961306396130120149183120149183
AMPS, Fuse4570110701208012517580125175150200250150200250
Cooling Water, 20°C GPM-363636936969126912
  • Humidity capability: 20% to 98% RH in the temperature range of +20°C to +85°C as limited by a 3°C dewpoint.
  • Humidity not achieved simultaneously with altitude.
  • Vacuum performance: All chambers achieve altitude up to 100,000 feet (8.2 mm Hg) in 35 minutes. Optional features extend altitude capabilities to 150,000 feet (1 mm Hg) and 200,000 feet (0.17 mm Hg).
  • Combined temperature and altitude to 60,000 feet maximum.
  • Test data based on 24°C amb., sea level, 60 Hz. Performance slightly reduced on 50 Hz. Power req. based on std. vacuum pump

Applications

  • Altitude
  • Temperature / Humidity
  • Vacuum

Industries

  • Aerospace
  • Defense
  • Environmental Processing